
The
CMM accommodates large, heavy workpieces with measurement ranges of up to 3,000
millimeters in the X-axis. Source: Werth
Tactile
and multisensor coordinate measuring machines (CMMs) are no longer used
exclusively in the inspection lab. More and more, they are used directly in
production. This means that operators must sometimes measure at temperatures
that deviate greatly from calibration temperature. Measurement errors resulting
from this temperature differential are often underestimated and neglected and,
therefore, cause errors in measurement results.
In addition to other effects, such as thermally induced deformation of the
measuring device and changes in probe lengths, the thermally induced linear
measurement deviations, due to temperature of the workpiece and the scales,
have been found to be a significant source of errors. Thus, appropriate
corrective measures must be taken when installing measuring devices in
production.
Thermally Induced Measurement Deviation
Temperatures that deviate from 20 C cause the scales in the CMM, and the part to
be measured, to expand by different amounts because of their different
materials. The influence of different temperatures compounds this problem.
Because these expansions have the same direction in length- measuring machines,
the only effect is the difference in expansion of the physical scale and the
workpiece. The influence of thermal expansion is largely linear. A simple linear
correction in each measurement axis can be implemented at a low cost. This is
sufficient for small measured lengths.
Thermally induced change in
length is principally calculated as follows:
ΔL = α
• L
0
• Δt
ΔL = thermally induced
change in length
α
= thermal expansion coefficient
L
0
= Reference length; in practice, the
measured length
Δt = Temperature deviation Δt = t
– 20° C
Expansion
of the scale:
ΔL
M
= α
M
• L
0
•
Δt
M
Expansion
of the workpiece:
ΔL
W
= α
W
• L
0
•
Δt
W
The thermally induced length measurement
deviation (Fig. 1) is the difference between ΔL
M
and ΔL
W
:
ΔL = L
0
(α
W
• Δt
W
- α
M
• Δt
M
)
As seen in the equation, with the same temperature deviation Dt at the
workpiece and the scale assumed for the simplest case, only the difference in
expansion coefficients determines the measured length deviation DL.
On a summer day at 35 C on the production floor, when measuring a
100-millimeter component made of PVDC without correction, an additional
measurement error of 208 microns is expected (See Table 1). In practice,
slightly different deviations are expected because the scale temperature and
workpiece temperature are usually different. This does not change the fact,
however, that the deviations are unacceptably large.
Lengthwise expansions of the scale and workpiece can compensate for one another
if the scale and workpiece are made of the same material, such as
steel-to-steel. In practice, this is usually not the case, so the expansion
coefficients are different. One can conclude that an appropriate correction for
the thermal influence is needed.

The
CMM offers shop floor production efficiency for small measurement ranges. Source:
Werth
Integrated Temperature Compensation
A
new series of CMMs provides a solid design base, made of granite with
encapsulated bearings. The devices are modular and can therefore be equipped to
meet exact operator requirements. The equipment includes image-processing
sensors, touch probes and laser probes. The machines allow simple manual
measurements as well as CNC-controlled measurement with 2-D and 3-D CAD
data.
All CMMs in the new series are designed to meet requirements of production
floor use. They can calculate corrections based on expansion coefficients of
the device and workpiece, workpiece temperature and associated temperatures of
the scales. Sensors measure these temperatures, and thermal expansion of the
scales is calculated automatically.

Table
1 Temperature-dependent, systematic measured length deviation, without
correction, for the same temperature at the workpiece and scales.
This shows possible measurement errors associated with the thermally induced,
measured-length deviation DL for workpieces of various materials at 20 C, 25 C,
30 C and 35 C ambient temperature. The values apply to a measured length of 100
mm, and for scales made of steel. It is also assumed that the workpiece and the
scales are at the same temperature. Source: Werth
In
the same way, the ambient temperature is measured near the workpiece, or
optionally, a contact thermometer is used. For each measured value, the
length-dependent correction value (DL) is calculated according to the formula
and measurement results are corrected accordingly. This correction also applies
to the machine diagonals because corrections are made in all three axes of the
measurement volume. Thus, the projected components are
considered.
Requirements are the entry of a precise value for expansion coefficients of the
workpiece and the maintenance of spatial (DK/m) and time (DK/h) temperature
gradients, in accordance with the manufacturer’s specifications. If the
expansion coefficient of the workpiece is not precisely known, an estimate of
the value will reduce the measured length deviation. This applies especially to
large measurement areas.

Table 2 the remaining thermally induced
measurement uncertainty with temperature correction.
Without
correction at 35 C, for a 100-millimeter PVDC workpiece, there is a measurement
error of 208 microns. With correction, the temperature influence is reduced as
shown here. Source: Werth
Results of Temperature Correction
If
temperature compensation is necessary, one cannot assume there will be no more
temperature influences. Despite the correction for thermal influences, there is
a remaining measurement uncertainty that is lower when the expansion
coefficient of the workpiece is known and the temperature is measured more
precisely. (See Table 2.)
Limit deviation is defined as the maximum possible (estimated) deviation from
the true value. It is assumed to be distributed rectangularly.
According to international standards, the expanded measurement uncertainty is
the standard deviation multiplied by the expansion factor k=2. Therefore, it
applies to a confidence interval of 95%.
For the selected measurement example, it is assumed that the temperature can be
measured at a limit deviation of 0.5 K. The expansion coefficient of the
workpiece is known with a maximum deviation of 5%. A measurement uncertainty of
16 microns thus remains. Here, that is only 7.7% of the original error of 207.8
microns without correction.

The
thermally induced length measurement deviation is the difference between the expansion
of the scale and the expansion of the workpiece. Expansion of the scale and the
workpiece have the opposite effect. Source: Werth
Tips for Practical Application
Temperature
correction is more reliable and effective if the following actions are
taken:
- Allow as little temperature change as possible.
- Avoid air drafts.
- Enclose the devices if rapid changes in temperature
occur.
- Keep heat sources in constant operation.
- Install devices at least 1 meter from the walls.
- Thermally insulate the floor.
- Do not allow direct sunlight.
- The measured object should be at the ambient
temperature.
- Avoid touching parts with hands, or wear gloves.
- Ensure constant temperature during a measurement
cycle.
- Carry out a drift check by reviewing the reference coordinate system,
for long measurement cycles.

The
CMM accommodates measurement ranges greater than 400 millimeters in the X and
Y-axes. Source: Werth
CMMs
can be used successfully in production, even under difficult temperature
conditions, if the operator follows these steps and uses proper temperature
correction. This will result in significantly fewer errors. These results
could, under some conditions, be much greater than the production tolerances
that are actually being inspected.
Tech tips
- Operators
must sometimes measure at temperatures that deviate greatly from calibration
temperature.
- For measuring small lengths, a simple linear correction in each
measurement axis can be implemented at a
low cost.
- Lengthwise expansions of the scale and workpiece can compensate for
one another if the scale and workpiece are made of the same
material.
- New modular CMMs measure temperatures with sensors and use scales to
calculate thermal expansion automatically.